Rotary Microtome HM 355 S
MICROM International GmbH
Robert-Bosch-Str. 49
D- 69190 Walldorf
PART 5 MAINTENANCE
Annual routine maintenance
To secure section quality and functioning of the
microtome, it is recommended that a routine
maintenance be performed by a trained service
technician once a year.
Service contract
MICROM offers a service contract which
guarantees that your instrument is always in
perfect condition. For more information, please
contact the nearest MICROM sales office.
Exchange of fuses
The two mains fuses (fig. 26.3) are installed on
the rear side of the instrument beside the power
switch (fig. 26.1).
To exchange the two fuses, turn off the power
switch of the instrument and unplug the
instrument. Open the fuse holder (fig. 26.2) from
the left side by means of a flat screw driver. Pull
the fuses off the fuse holder and replace the fuses
with new ones.
Rating of fuses:
For power requirements: 220-240 V:
- 2 fuses 1,6 AT slow-blow (cat. no. 326670)
For power requirements: 100-120 V:
- 2 fuses 3,2 AT slow-blow (cat. no. 326660)
Fig. 26
Exchange of fuse for the motor supply
The fuse for the motor supply (fig. 26.4) is placed
on the rear side of the instrument on the right side
of the mains plug unit and secures the supply
voltage of the cutting drive.
To exchange the two fuses, turn off the power
switch of the instrument and unplug the
instrument.
Press the fuse holder cap by means of a
screwdriver and turn it to the right side.
Remove the fuse and replace it with a new one.
Insert the cap with the fuse into the fuse holder
and lock the bajonet catch.
Rating of fuse:
- 1 fuse 5,0 AT, slow-blow (cat. no. 326350)